Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket

Noll, Veronika ; Rinderknecht, Stephan ; Beckerle, Philipp (2018)
Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket.
In: Sensors, 2018, 18 (7)
Artikel, Zweitveröffentlichung, Verlagsversion

WarnungEs ist eine neuere Version dieses Eintrags verfügbar.

Kurzbeschreibung (Abstract)

A sensor system for measuring the relative movement between prosthetic socket and residual limb based on a 2D-motion sensor is introduced and thoroughly tested experimentally. The quantitative analysis of test rig evaluation is used to identify advantageous sensor settings and liner configurations. Considering these favorable settings, sensor functionality is quantified to errrel=0.52±1.78%. Advancing to convex measurement surfaces, the sensor shows absolute errors of errabs≤1 mm in an observable measurement scenario. The feasibility of measuring gait-induced relative movement with the proposed 2D-motion sensor is shown via a biomechanical plausibility study. Overall, the findings suggest that the proposed sensor system is suitable for investigating the relative movement between residual limb and prosthetic socket in dynamic gait situations

Typ des Eintrags: Artikel
Erschienen: 2018
Autor(en): Noll, Veronika ; Rinderknecht, Stephan ; Beckerle, Philipp
Art des Eintrags: Zweitveröffentlichung
Titel: Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket
Sprache: Englisch
Publikationsjahr: 2018
Ort: Darmstadt
Publikationsdatum der Erstveröffentlichung: 2018
Titel der Zeitschrift, Zeitung oder Schriftenreihe: Sensors
Jahrgang/Volume einer Zeitschrift: 18
(Heft-)Nummer: 7
URL / URN: https://tuprints.ulb.tu-darmstadt.de/7652
Zugehörige Links:
Herkunft: Zweitveröffentlichung aus gefördertem Golden Open Access
Kurzbeschreibung (Abstract):

A sensor system for measuring the relative movement between prosthetic socket and residual limb based on a 2D-motion sensor is introduced and thoroughly tested experimentally. The quantitative analysis of test rig evaluation is used to identify advantageous sensor settings and liner configurations. Considering these favorable settings, sensor functionality is quantified to errrel=0.52±1.78%. Advancing to convex measurement surfaces, the sensor shows absolute errors of errabs≤1 mm in an observable measurement scenario. The feasibility of measuring gait-induced relative movement with the proposed 2D-motion sensor is shown via a biomechanical plausibility study. Overall, the findings suggest that the proposed sensor system is suitable for investigating the relative movement between residual limb and prosthetic socket in dynamic gait situations

Status: Verlagsversion
URN: urn:nbn:de:tuda-tuprints-76520
Sachgruppe der Dewey Dezimalklassifikatin (DDC): 600 Technik, Medizin, angewandte Wissenschaften > 600 Technik
Fachbereich(e)/-gebiet(e): 16 Fachbereich Maschinenbau
16 Fachbereich Maschinenbau > Institut für Mechatronische Systeme im Maschinenbau (IMS)
Hinterlegungsdatum: 31 Jul 2018 13:26
Letzte Änderung: 25 Mai 2023 08:56