Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket

Noll, Veronika and Rinderknecht, Stephan and Beckerle, Philipp (2018):
Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket.
In: Sensors, p. 2170, 18, (7), ISSN 1424-8220,
DOI: 10.3390/s18072170,
[Online-Edition: https://doi.org/10.3390/s18072170],
[Article]

Abstract

A sensor system for measuring the relative movement between prosthetic socket and residual limb based on a 2D-motion sensor is introduced and thoroughly tested experimentally. The quantitative analysis of test rig evaluation is used to identify advantageous sensor settings and liner configurations. Considering these favorable settings, sensor functionality is quantified to errrel=0.52±1.78%. Advancing to convex measurement surfaces, the sensor shows absolute errors of errabs≤1 mm in an observable measurement scenario. The feasibility of measuring gait-induced relative movement with the proposed 2D-motion sensor is shown via a biomechanical plausibility study. Overall, the findings suggest that the proposed sensor system is suitable for investigating the relative movement between residual limb and prosthetic socket in dynamic gait situations

Item Type: Article
Erschienen: 2018
Creators: Noll, Veronika and Rinderknecht, Stephan and Beckerle, Philipp
Title: Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket
Language: English
Abstract:

A sensor system for measuring the relative movement between prosthetic socket and residual limb based on a 2D-motion sensor is introduced and thoroughly tested experimentally. The quantitative analysis of test rig evaluation is used to identify advantageous sensor settings and liner configurations. Considering these favorable settings, sensor functionality is quantified to errrel=0.52±1.78%. Advancing to convex measurement surfaces, the sensor shows absolute errors of errabs≤1 mm in an observable measurement scenario. The feasibility of measuring gait-induced relative movement with the proposed 2D-motion sensor is shown via a biomechanical plausibility study. Overall, the findings suggest that the proposed sensor system is suitable for investigating the relative movement between residual limb and prosthetic socket in dynamic gait situations

Journal or Publication Title: Sensors
Volume: 18
Number: 7
Divisions: 16 Department of Mechanical Engineering
16 Department of Mechanical Engineering > Institute for Mechatronic Systems in Mechanical Engineering (IMS)
Interdisziplinäre Forschungsprojekte
Interdisziplinäre Forschungsprojekte > Forschungsprojekte Prothetik
Date Deposited: 09 Jul 2018 07:02
DOI: 10.3390/s18072170
Official URL: https://doi.org/10.3390/s18072170
URN: urn:nbn:de:tuda-tuprints-76520
Export:
Suche nach Titel in: TUfind oder in Google

Optionen (nur für Redakteure)

View Item View Item