Noll, Veronika ; Rinderknecht, Stephan ; Beckerle, Philipp (2018)
Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket.
In: Sensors, 18 (7)
doi: 10.3390/s18072170
Artikel, Bibliographie
Dies ist die neueste Version dieses Eintrags.
Kurzbeschreibung (Abstract)
A sensor system for measuring the relative movement between prosthetic socket and residual limb based on a 2D-motion sensor is introduced and thoroughly tested experimentally. The quantitative analysis of test rig evaluation is used to identify advantageous sensor settings and liner configurations. Considering these favorable settings, sensor functionality is quantified to errrel=0.52±1.78%. Advancing to convex measurement surfaces, the sensor shows absolute errors of errabs≤1 mm in an observable measurement scenario. The feasibility of measuring gait-induced relative movement with the proposed 2D-motion sensor is shown via a biomechanical plausibility study. Overall, the findings suggest that the proposed sensor system is suitable for investigating the relative movement between residual limb and prosthetic socket in dynamic gait situations
Typ des Eintrags: | Artikel |
---|---|
Erschienen: | 2018 |
Autor(en): | Noll, Veronika ; Rinderknecht, Stephan ; Beckerle, Philipp |
Art des Eintrags: | Bibliographie |
Titel: | Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket |
Sprache: | Englisch |
Publikationsjahr: | 6 Juli 2018 |
Titel der Zeitschrift, Zeitung oder Schriftenreihe: | Sensors |
Jahrgang/Volume einer Zeitschrift: | 18 |
(Heft-)Nummer: | 7 |
DOI: | 10.3390/s18072170 |
URL / URN: | https://doi.org/10.3390/s18072170 |
Zugehörige Links: | |
Kurzbeschreibung (Abstract): | A sensor system for measuring the relative movement between prosthetic socket and residual limb based on a 2D-motion sensor is introduced and thoroughly tested experimentally. The quantitative analysis of test rig evaluation is used to identify advantageous sensor settings and liner configurations. Considering these favorable settings, sensor functionality is quantified to errrel=0.52±1.78%. Advancing to convex measurement surfaces, the sensor shows absolute errors of errabs≤1 mm in an observable measurement scenario. The feasibility of measuring gait-induced relative movement with the proposed 2D-motion sensor is shown via a biomechanical plausibility study. Overall, the findings suggest that the proposed sensor system is suitable for investigating the relative movement between residual limb and prosthetic socket in dynamic gait situations |
Sachgruppe der Dewey Dezimalklassifikatin (DDC): | 600 Technik, Medizin, angewandte Wissenschaften > 600 Technik |
Fachbereich(e)/-gebiet(e): | 16 Fachbereich Maschinenbau 16 Fachbereich Maschinenbau > Institut für Mechatronische Systeme im Maschinenbau (IMS) Interdisziplinäre Forschungsprojekte Interdisziplinäre Forschungsprojekte > Forschungsprojekte Prothetik |
Hinterlegungsdatum: | 09 Jul 2018 07:02 |
Letzte Änderung: | 03 Jul 2024 02:30 |
PPN: | |
Export: | |
Suche nach Titel in: | TUfind oder in Google |
Verfügbare Versionen dieses Eintrags
-
Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket. (deposited 31 Jul 2018 13:26)
- Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket. (deposited 09 Jul 2018 07:02) [Gegenwärtig angezeigt]
Frage zum Eintrag |
Optionen (nur für Redakteure)
Redaktionelle Details anzeigen |