Masalovich, S. ; Ioffe, A. ; Schlapp, Michael ; Brückel, T. ; Seggern, Heinz von (2005)
Optimization of a neutron image plate detector with low gamma-sensitivity.
In: Nuclear instruments & methods in physics research / A, 539 (1-2)
Artikel, Bibliographie
Kurzbeschreibung (Abstract)
Recent experimental studies on new materials (storage phosphors and neutron converter) to be used in a low γγ-sensitive neutron image plate detector have been performed at the Research Center Jülich and Darmstadt University of Technology. It was shown that a mixture LiF+KCl:Eu2+ (or KBr:Eu2+) features a higher neutron/gamma signal ratio compared to BaFBr:Eu2+mixed with Gd as in commercially available neutron image plates. However, the smaller neutron capture cross-section of 6Li in comparison with Gd requires the use of significantly thicker recording layers. This article presents the results of numerical estimations for the performance of such a detector. The use of image plates with a thick recording layer and a regular cell structure allowing for the increase in spatial resolution (pixelated image plate) is discussed.
Typ des Eintrags: | Artikel |
---|---|
Erschienen: | 2005 |
Autor(en): | Masalovich, S. ; Ioffe, A. ; Schlapp, Michael ; Brückel, T. ; Seggern, Heinz von |
Art des Eintrags: | Bibliographie |
Titel: | Optimization of a neutron image plate detector with low gamma-sensitivity |
Sprache: | Englisch |
Publikationsjahr: | 21 Februar 2005 |
Verlag: | Elsevier Science Publishing |
Titel der Zeitschrift, Zeitung oder Schriftenreihe: | Nuclear instruments & methods in physics research / A |
Jahrgang/Volume einer Zeitschrift: | 539 |
(Heft-)Nummer: | 1-2 |
URL / URN: | http://www.sciencedirect.com/science/article/pii/S0168900204... |
Kurzbeschreibung (Abstract): | Recent experimental studies on new materials (storage phosphors and neutron converter) to be used in a low γγ-sensitive neutron image plate detector have been performed at the Research Center Jülich and Darmstadt University of Technology. It was shown that a mixture LiF+KCl:Eu2+ (or KBr:Eu2+) features a higher neutron/gamma signal ratio compared to BaFBr:Eu2+mixed with Gd as in commercially available neutron image plates. However, the smaller neutron capture cross-section of 6Li in comparison with Gd requires the use of significantly thicker recording layers. This article presents the results of numerical estimations for the performance of such a detector. The use of image plates with a thick recording layer and a regular cell structure allowing for the increase in spatial resolution (pixelated image plate) is discussed. |
Freie Schlagworte: | Neutron detector, Image plate, Detector efficiency, Pixelated detector |
Zusätzliche Informationen: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment |
Fachbereich(e)/-gebiet(e): | 11 Fachbereich Material- und Geowissenschaften 11 Fachbereich Material- und Geowissenschaften > Materialwissenschaft 11 Fachbereich Material- und Geowissenschaften > Materialwissenschaft > Elektronische Materialeigenschaften |
Hinterlegungsdatum: | 20 Nov 2008 08:21 |
Letzte Änderung: | 20 Feb 2020 13:24 |
PPN: | |
Sponsoren: | The present study was supported by the EU in the frame of the TECHNI project (Contract No. HPRI-CT-1999-50005). |
Export: | |