TU Darmstadt / ULB / TUbiblio

A 3D-Elastography-Guided System for Laparoscopic Partial Nephrectomies

Stolka, Philipp J. ; Keil, Matthias ; Sakas, Georgios ; McVeigh, Elliot R. ; Allaf, Mohamad E. ; Taylor, Russell H. ; Boctor, Emad (2010)
A 3D-Elastography-Guided System for Laparoscopic Partial Nephrectomies.
Medical Imaging 2010: Visualization, Image-Guided Procedures, and Modeling. Part One.
Konferenzveröffentlichung, Bibliographie

Kurzbeschreibung (Abstract)

We present an image-guided intervention system based on tracked 3D elasticity imaging (EI) to provide a novel interventional modality for registration with pre-operative CT. The system can be integrated in both laparoscopic and robotic partial nephrectomies scenarios, where this new use of EI makes exact intra-operative execution of pre-operative planning possible. Quick acquisition and registration of 3D-B-Mode and 3D-EI volume data allows intra-operative registration with CT and thus with pre-defined target and critical regions (e.g. tumors and vasculature). Their real-time location information is then overlaid onto a tracked endoscopic video stream to help the surgeon avoid vessel damage and still completely resect tumors including safety boundaries. The presented system promises to increase the success rate for partial nephrectomies and potentially for a wide range of other laparoscopic and robotic soft tissue interventions. This is enabled by the three components of robust real-time elastography, fast 3D-EI/CT registration, and intra-operative tracking. With high quality, robust strain imaging (through a combination of parallelized 2D-EI, optimal frame pair selection, and optimized palpation motions), kidney tumors that were previously unregistrable or sometimes even considered isoechoic with conventional B-mode ultrasound can now be imaged reliably in interventional settings. Furthermore, this allows the transformation of planning CT data of kidney ROIs to the intra-operative setting with a markerless mutual-information-based registration, using EM sensors for intraoperative motion tracking. Overall, we present a complete procedure and its development, including new phantom models - both ex vivo and synthetic - to validate image-guided technology and training, tracked elasticity imaging, real-time EI frame selection, registration of CT with EI, and finally a real-time, distributed software architecture. Together, the system allows the surgeon to concentrate on intervention completion with less time pressure.

Typ des Eintrags: Konferenzveröffentlichung
Erschienen: 2010
Autor(en): Stolka, Philipp J. ; Keil, Matthias ; Sakas, Georgios ; McVeigh, Elliot R. ; Allaf, Mohamad E. ; Taylor, Russell H. ; Boctor, Emad
Art des Eintrags: Bibliographie
Titel: A 3D-Elastography-Guided System for Laparoscopic Partial Nephrectomies
Sprache: Englisch
Publikationsjahr: 2010
Verlag: SPIE Press, Bellingham
Reihe: Proceedings of SPIE; 7625
Veranstaltungstitel: Medical Imaging 2010: Visualization, Image-Guided Procedures, and Modeling. Part One
Kurzbeschreibung (Abstract):

We present an image-guided intervention system based on tracked 3D elasticity imaging (EI) to provide a novel interventional modality for registration with pre-operative CT. The system can be integrated in both laparoscopic and robotic partial nephrectomies scenarios, where this new use of EI makes exact intra-operative execution of pre-operative planning possible. Quick acquisition and registration of 3D-B-Mode and 3D-EI volume data allows intra-operative registration with CT and thus with pre-defined target and critical regions (e.g. tumors and vasculature). Their real-time location information is then overlaid onto a tracked endoscopic video stream to help the surgeon avoid vessel damage and still completely resect tumors including safety boundaries. The presented system promises to increase the success rate for partial nephrectomies and potentially for a wide range of other laparoscopic and robotic soft tissue interventions. This is enabled by the three components of robust real-time elastography, fast 3D-EI/CT registration, and intra-operative tracking. With high quality, robust strain imaging (through a combination of parallelized 2D-EI, optimal frame pair selection, and optimized palpation motions), kidney tumors that were previously unregistrable or sometimes even considered isoechoic with conventional B-mode ultrasound can now be imaged reliably in interventional settings. Furthermore, this allows the transformation of planning CT data of kidney ROIs to the intra-operative setting with a markerless mutual-information-based registration, using EM sensors for intraoperative motion tracking. Overall, we present a complete procedure and its development, including new phantom models - both ex vivo and synthetic - to validate image-guided technology and training, tracked elasticity imaging, real-time EI frame selection, registration of CT with EI, and finally a real-time, distributed software architecture. Together, the system allows the surgeon to concentrate on intervention completion with less time pressure.

Freie Schlagworte: Ultrasound, Navigation, Registration, Distributed applications
Fachbereich(e)/-gebiet(e): 20 Fachbereich Informatik
20 Fachbereich Informatik > Graphisch-Interaktive Systeme
Hinterlegungsdatum: 12 Nov 2018 11:16
Letzte Änderung: 12 Nov 2018 11:16
PPN:
Export:
Suche nach Titel in: TUfind oder in Google
Frage zum Eintrag Frage zum Eintrag

Optionen (nur für Redakteure)
Redaktionelle Details anzeigen Redaktionelle Details anzeigen