TU Darmstadt / ULB / TUbiblio

Flow and mass transfer prediction in anisotropic TPMS-structures as extracorporeal oxygenator membranes using reduced order modeling

Barbian, Kai P. ; Hirschwald, Lukas T. ; Linkhorst, John ; Neidlin, Michael ; Steinseifer, Ulrich ; Wessling, Matthias ; Wiegmann, Bettina ; Jansen, Sebastian V. (2024)
Flow and mass transfer prediction in anisotropic TPMS-structures as extracorporeal oxygenator membranes using reduced order modeling.
In: Journal of Membrane Science, 690
doi: 10.1016/j.memsci.2023.122160
Article, Bibliographie

Abstract

Currently, hollow fiber membranes are the standard technology for extracorporeal membrane oxygenators. Apart from the inevitable contact of the circulating blood with the artificial material, a suboptimal flow distribution within the oxygenator favors thrombus formation which leads to a rapid loss of gas exchange capacity. The current advancement in additive manufacturing allows the design of three-dimensional membrane-structures, based on triply periodic minimal surfaces (TPMS). One of their unique advantages is local geometry variation to manipulate the flow distribution. But how this anisotropy influences the overall device performance is non-trivial and requires numerical simulation. The aim of this study was to develop a reduced order model (ROM) that is able to efficiently predict three-dimensional flow distribution and gas transfer inside TPMS-structures. We performed a parametric study using a validated micro scale computational fluid dynamics (CFD) model. Afterwards, two different modeling approaches of Sherwood-correlations and artificial neural networks (NN) were compared to characterize flow and mass transfer from the simulated data. To create the ROM, the NN modeling strategy was then implemented into a porous medium CFD model. The developed ROM was also validated. Finally, an anisotropic TPMS-membrane-structure was compared numerically with an isotropic predicate. The NN fitting strategy showed superior accuracy over the Sherwood-correlations for characterizing mass transfer in TPMS-structures. With the ROM, the gas transfer rates of oxygen and carbon dioxide from the micro CFD model could be predicted within relative root mean squared errors (RRMSE) of 2.7% and 5.1%. The pressure drop in the experiments was predicted with an RRMSE accuracy of 7.6%. In comparison with the isotropic TPMS-structure, the anisotropic structure showed a homogenized flow distribution and increased gas transfer rates of 4% to 5% for oxygen and 2.3% to 7.4% for carbon dioxide at the simulated flow rates.

Item Type: Article
Erschienen: 2024
Creators: Barbian, Kai P. ; Hirschwald, Lukas T. ; Linkhorst, John ; Neidlin, Michael ; Steinseifer, Ulrich ; Wessling, Matthias ; Wiegmann, Bettina ; Jansen, Sebastian V.
Type of entry: Bibliographie
Title: Flow and mass transfer prediction in anisotropic TPMS-structures as extracorporeal oxygenator membranes using reduced order modeling
Language: English
Date: January 2024
Publisher: Elsevier
Journal or Publication Title: Journal of Membrane Science
Volume of the journal: 690
DOI: 10.1016/j.memsci.2023.122160
Abstract:

Currently, hollow fiber membranes are the standard technology for extracorporeal membrane oxygenators. Apart from the inevitable contact of the circulating blood with the artificial material, a suboptimal flow distribution within the oxygenator favors thrombus formation which leads to a rapid loss of gas exchange capacity. The current advancement in additive manufacturing allows the design of three-dimensional membrane-structures, based on triply periodic minimal surfaces (TPMS). One of their unique advantages is local geometry variation to manipulate the flow distribution. But how this anisotropy influences the overall device performance is non-trivial and requires numerical simulation. The aim of this study was to develop a reduced order model (ROM) that is able to efficiently predict three-dimensional flow distribution and gas transfer inside TPMS-structures. We performed a parametric study using a validated micro scale computational fluid dynamics (CFD) model. Afterwards, two different modeling approaches of Sherwood-correlations and artificial neural networks (NN) were compared to characterize flow and mass transfer from the simulated data. To create the ROM, the NN modeling strategy was then implemented into a porous medium CFD model. The developed ROM was also validated. Finally, an anisotropic TPMS-membrane-structure was compared numerically with an isotropic predicate. The NN fitting strategy showed superior accuracy over the Sherwood-correlations for characterizing mass transfer in TPMS-structures. With the ROM, the gas transfer rates of oxygen and carbon dioxide from the micro CFD model could be predicted within relative root mean squared errors (RRMSE) of 2.7% and 5.1%. The pressure drop in the experiments was predicted with an RRMSE accuracy of 7.6%. In comparison with the isotropic TPMS-structure, the anisotropic structure showed a homogenized flow distribution and increased gas transfer rates of 4% to 5% for oxygen and 2.3% to 7.4% for carbon dioxide at the simulated flow rates.

Uncontrolled Keywords: anisotropic, gas transfer simulation, membrane oxygenator, reduced order modeling,TPMS
Identification Number: Artikel-ID: 122160
Divisions: 16 Department of Mechanical Engineering
16 Department of Mechanical Engineering > Chair for Process Engineering of Electrochemical Systems
Date Deposited: 24 Jul 2024 07:06
Last Modified: 24 Jul 2024 07:06
PPN:
Export:
Suche nach Titel in: TUfind oder in Google
Send an inquiry Send an inquiry

Options (only for editors)
Show editorial Details Show editorial Details