On the Balance between Albumin Loss and Removal of Middle Molecules in Dialyzers

Hagemann, Franziska ; Linkhorst, John ; Roth, Hannah ; Wessling, Matthias (2023)
On the Balance between Albumin Loss and Removal of Middle Molecules in Dialyzers.
In: Journal of Membrane Science Letters, 3 (1)
doi: 10.1016/j.memlet.2023.100044
Article, Bibliographie

Abstract

Tuning the pore size distribution of hemodialysis membranes is essential for the membrane's selectivity and significantly affects the quality of the dialysis treatment. Tailoring the membrane's molecular weight cut-off appropriately balances the removal of middle-molecular-weight uremic toxins while avoiding albumin loss. This undesirable albumin loss is a potential side effect and concern for clinical use when aiming at increased removal of middle molecular weight molecules (middle molecules). It is hypothesized that control of the position of a narrow pore size distribution allows middle molecule removal while simultaneously counteracting the unwanted albumin loss. This study presents a comprehensive ex-vivo methodology and novel data on the balance of the clearance of middle molecules and albumin loss at different blood and dialysate flow rates using novel dialyzers. The outcomes hold significance for dialysis therapy, while the insights acquired have broader implications for the selectivity of ultrafiltration membranes. The Theranova dialyzer shows the largest clearance for small-middle molecules. Phylther stands out with higher removal of the middle molecule YKL-40 than the other dialyzers but exhibits a significant albumin loss. Theranova demonstrates the best compromise between low albumin loss and good clearances of middle molecules.

Item Type: Article
Erschienen: 2023
Creators: Hagemann, Franziska ; Linkhorst, John ; Roth, Hannah ; Wessling, Matthias
Type of entry: Bibliographie
Title: On the Balance between Albumin Loss and Removal of Middle Molecules in Dialyzers
Language: English
Date: 2023
Publisher: Elsevier
Journal or Publication Title: Journal of Membrane Science Letters
Volume of the journal: 3
Issue Number: 1
DOI: 10.1016/j.memlet.2023.100044
Abstract:

Tuning the pore size distribution of hemodialysis membranes is essential for the membrane's selectivity and significantly affects the quality of the dialysis treatment. Tailoring the membrane's molecular weight cut-off appropriately balances the removal of middle-molecular-weight uremic toxins while avoiding albumin loss. This undesirable albumin loss is a potential side effect and concern for clinical use when aiming at increased removal of middle molecular weight molecules (middle molecules). It is hypothesized that control of the position of a narrow pore size distribution allows middle molecule removal while simultaneously counteracting the unwanted albumin loss. This study presents a comprehensive ex-vivo methodology and novel data on the balance of the clearance of middle molecules and albumin loss at different blood and dialysate flow rates using novel dialyzers. The outcomes hold significance for dialysis therapy, while the insights acquired have broader implications for the selectivity of ultrafiltration membranes. The Theranova dialyzer shows the largest clearance for small-middle molecules. Phylther stands out with higher removal of the middle molecule YKL-40 than the other dialyzers but exhibits a significant albumin loss. Theranova demonstrates the best compromise between low albumin loss and good clearances of middle molecules.

Uncontrolled Keywords: Albumin loss, Ex-vivo dialysis, Membrane module characterization, Middle molecule removal
Divisions: 16 Department of Mechanical Engineering
16 Department of Mechanical Engineering > Chair for Process Engineering of Electrochemical Systems
Date Deposited: 13 Sep 2023 11:13
Last Modified: 13 Sep 2023 11:13
PPN:
Export:
Suche nach Titel in: TUfind oder in Google
Send an inquiry Send an inquiry

Options (only for editors)
Show editorial Details Show editorial Details